Impac IS 12-TSP and IGA 12-TSP
Fully Digital, Extremely Precise Infrared Thermometer, 200 to 3000°C
Advanced Energy's Impac IS 12-TSP and IGA 12-TSP are fully digital, extremely precise transfer standard pyrometers for exact inspection of calibration sources.
- Temperature Ranges: Between 200 and 3000°C
- Analog Output: 0 to 20 mA or 4 to 20 mA, DC, switchable, load max. 500 Ohm
- Extremely fast exposure time
- Long-term stability
- Thermostatically-controlled detectors
- Overview
- Technical Specs
- Documents
- Accessories
- Service & Support
- Distributors
Overview
The Impac® IS 12-TSP and IGA 12-TSP are extremely precise transfer standard pyrometers that can inspect calibration sources. The detectors in TS pyrometers are thermostatically controlled to achieve high precision. This means measurements are, to a large extent, independent of surrounding temperature variations and a resolution of 0.01°C can be achieved.
The IS 12-TSP or IGA 12-TSP are available in several temperature ranges between 200 and 3000°C. You can also select from various spectral ranges. This enables use with calibration sources that check pyrometers with silicon or indium gallium arsenide detectors.
Benefits
- Accurately inspect your calibration sources
- Use with calibration sources with which pyrometers with silicon or indium gallium arsenide detectors can be checked
- Achieve high precision
- Ensure that temperature values, as stipulated by national institutes, can be transferred to your own calibration sources to guarantee traceability to the ITS90 international temperature scale
- Guarantee adherence to high technical specifications
Features
- Temperature ranges between 200 and 3000°C
- Resolution of only 0.01°C
- Built-in, five-digit LED display
- Digital interface
- Focusable optics
Select specifications are shown below. Please see product data sheet for additional specifications.
-
Temperature Ranges
IS 12-TSP
530 to 1900°C (986 to 3452°F)
600 to 2520°C (1112 to 4568°F)
600 to 3000°C (1112 to 5432°F)
850 to 2520°C (1562 to 4568°F)IGA 12-TSP
200 to 1020°C (392 to 1868°F)
250 to 1400°C (482 to 2552°F)
300 to 1700°C (572 to 3092°F)Any range adjustable within the temperature range, minimum span 51°C
-
Measurement
Spectral Ranges
Measurement Uncertainty
Below 1500°C: 0.15% of measured value in °C + 1°C
Above 1500°C: 0.25% of measured value in °C
Above 2700°C: 0.35% of measured value in °C
ε = 1, t90 = 1 s, Tamb. = 23°CRepeatability
1°C
ε = 1, t90 = 1 s, Tamb. = 23°CExposure Time t90
< 1 ms (with dynamical adaptation at low signal levels), factory setting 1 s, adjustable up to 10 s
Emissivity ε
0.100 to 1.000 in 1⁄1000 steps
-
Optical
Sighting
Built-in parallax free through-lens view finder and laser targeting light
IS 12-TSP with temperature range 850 to 2520°C: only with view finderOptics
Three focusable optics
Distance Ratio
IS 12-TSP: 400:1 (0.7 mm)
IGA 12-TSP: 250:1 (1.1 mm) -
Interface and Communication
Display
Built-in, five-digit LED display, additional function LEDs
Parameters
Adjustable at the instrument or via serial interface: Emissivity ε, exposure time t90, clear time for maximum value storage tclear, temperature subrange, analog output 0 to 20 or 4 to 20 mA, switch points for limit switches, temperature display in °C / °F, interface RS232 or RS485, address, baud rate, test current output
Additionally adjustable (only via interface): keyboard lock, recalibration (with special software)
Analog Output
Linear 0 to 20 mA or 4 to 20 mA, DC, switchable; load maximum 500 Ohm
Digital Interface
Switchable at the pyrometer, RS232 or RS485 (addressable), half duplex; baud rate 2.4 up to 115 kBd
Limit Switches
Two relay outputs (change-over contacts), switch power max 30 W (Imax: 1 A, Umax: 60 VDC)
-
Environmental
Protection Class
IP65 (according to DIN 40 050)
Ambient Temperature
0 to 60°C (32 to 140°F) at the housing
Storage Temperature
-20 to 70°C (-4 to 158°F)
Relative Humidity
Non-condensing conditions
Weight
2.2 kg (~2.20 lb)
CE Label
According to EU directives about electromagnetic immunity
-
Electrical
Power Supply
24 VDC (15 to 40 VDC), or 24 VAC (12 to 30 VAC), 48 to 62 Hz
Power Consumption
Maximum 14 W
Isolation
Power supply, digital interface, analog output are galvanically isolated against each other and housing
Specifications are subject to change without notice. Not all options and accessories are available for all models and configurations. Dimensions listed generally refer to standard configurations and module bodies only. Please confirm technical specifications and customizations with a sales representative. Advanced Energy is not responsible for errors or omissions.
Data Sheets
Brochures
Manuals
Application Notes
Accessories
Global Support & Services
Maximize fab productivity and capital equipment ROI with world-class support through each product lifecycle stage — from startup to long-term operation. Our service offerings are based on more than three decades of precision power and applications expertise.
Applications Support & Consulting
Accelerate tool installations and process development programs, or address specific applications concerns with Advanced Energy's dedicated applications consultants and engineering staff. Benefit from on-site reviews, analyses, and consultations.
Warranties
Decrease variability in maintenance costs and provide additional cost protection. Advanced Energy offers whole-box extended warranties.
Product Support
Call
U.S.: +1 800 446 9167
Asia: +86 29 8874 1895
China: 400 8899 130
EMEA: +44 800 032 1546
U.K.: 08000 321 546
Email
Thin Film Support
Thermal Product Support
Other Technical Support
Find an AE Representative or Distributor
To find distributors and representatives who sell AE products in your area, select your location and vendor type (optional).